Technical Specifications Radiation Probe DLMon
Intelligent Radiation Detector: Type SON 16
Measured value: | Photon dose rate equivalent respectively |
| Ambient dose rate equivalent |
| Detector: | Geiger-Mueller-Tube / Type 70004 |
| Measurement range: | 0.1 µSv/h* ... 500 µSv/h |
| *( Measuring period 200 s, measuring uncertainity < 3 % ) |
| Energy range: | 35 keV ... 1,3 MeV |
| Collimator: | Material: lead (Pb) |
| Aperture angle: ± 30 ° |
| Electronics: | - Micro-controller with internal memory
- 128-kByte-FLASH-EEPROM
- Data transfer by a RS-485 interface
- Calculation of dose rate from counting rate
- Correction of detector dead time and background counting rate
- Data memory (up to 9088 values) including date and time
in a non-volatile cycling memory, even at a breakdown of the PC - Acquisition parameters (cquisition time, number of cycles)
can be defined - Continuous self-monitoring of all internal and external
operating voltages, functional tests of the detector, of the interface and memory
|
|
|
Intelligent Radiation Detector: Type SON 20

|
|
| Measurement values: | Photon dose rate equivalent respectively |
| Ambient dose rate equivalent |
Detector: | - Geiger-Mueller-Tube / Type 70013E (energy compensated for the photon equivalent dose rate)
- Geiger-Mueller-Tube / Type 70013A (energy compensated for the ambient equivalent dose rate
|
| Measurement range: | 0.05 µSv/h* ... 200 µSv/h |
| *( Measuring period 200 s, measuring uncertainity < 3 % ) |
| Energy range: | 35 keV ... 1,3 MeV |
| Collimator: | Without |
| Electronics: | - Micro-controller with internal memory
- 128-kByte-FLASH-EEPROM
- Data transfer by an RS-485 interface
- Calculation of dose rate from counting rate
- Correction of detector dead time and background counting rate
- Data memory (up to 9088 values) including date and time
in a non-volatile cycling memory, even at a breakdown of the PC - Acquisition parameters (acquisition time, number of cycles)
can be defined - Continuous self-monitoring of all internal and external
operating voltages, functional tests of the detector, of the interface and memory
|
Options:
| - Lead collimator
- Potential free display for alarming threshold overruns and disturbances, every output can withstand 24 volt and 0.5 ampere
|

Intelligent Exhaust Air Sensor: Type AL 2000
System description:
The AL 2000 system is for monitoring of radioactive gases in the exhaust air.
Within the all-side shielding is a detection system that detects the background effect of the measuring station as well as radioactive air samples. A pump transports a part of the exhaust air with a continuous flow through the measuring station. All detectors are sending impulses, whose handling does an electronically counting-mechanism. The measurment values are relayed to a PC over a bus-system. This PC performs the archiving and data-processing, controls and monitors all processes and parameters at one time.
|
|
| Dimensions (H x L x W): | 560 mm x 510 mm x 630 mm |
| Weight: | 600 kg |
| Measuring chamber: | Detector: 3 XE-detector RXD 1000 |
| Window: 1000 cm², 15 µm Ti-foil |
| Background counting rate: 17,5 cps |
| Volume: 5 litres, intake and outtake each with a |
| gas distribution chamber |
| Shielding: | 50 mm thick wall of lead (all-side) |
| Functional control: | by a test radiator |
| Electronics: | - Micro-controller with internal memory
- 128-kByte-FLASH-EEPROM
- Data transfer by an RS-485 interface
- Calculation of volume activity from counting rate
- Correction of detector dead time and background counting rate
- Data memory (up to 9088 values) including date and time
in a non-volatile cycling memory, even at a breakdown of the PC - Acquisition parameters (acquisition time, number of cycles)
can be defined - Continuous self-monitoring of all internal and external
operating voltages, functional tests of the detector, of the interface and memory
|