Technical Specifications Polimaster PM1703MO-1
| Indicating range for dose rate equivalent: Measurement range for dose rate equivalent: |
0.01 µSv/h .. 9.99Sv/h 0.1 µSv/h .. 9.99Sv/h |
| Tolerable intrinsic error for dose rate equivalent: |
±(20 +K1/H +K2*H)%, H means dose rate equivalent in mSv/h K1 = 0.0025mSv/h K2 = 1/(0.002mSv/h) |
| Indicating range for dose equivalent: Measurement range for dose equivalent: |
0.01 µSv .. 9.99Sv 0.1 µSv .. 9.99Sv |
| Sensitivity for gamma radiation in search mode: |
100.0/(s * µSv/h) for Am-241 85.0/(s * µSv/h) for Cs-137 |
| Energy range in search mode: | 0.033 .. 3.0 MeV |
| Response time at rising of a Cs-137-radiation source for about 1µSv/h within 0.5 seconds: | max. 4 seconds |
| Types of alarms: | Sound, vibration, visual |
| Power supply: |
1x AA-battery LR6 1.5 V Alkaline or AA-NiMH-battery with min. 2000mAh |
| Battery lifetime: | Up to 1000 hours |
| Capacity of the non-volatile memory: | Up to 2000 data sets |
| Operating temperature range: | -30 .. 50 °C |
| Degree of protection: | IP65, water proofed and dust proofed |
| Max. height for break-proof in case of falling onto a concrete area: | 0.7m |
| Dimensions: |
72 x 32 x 87 mm |
| Weight (with batteries): | max. 250g |