Technical Specifications Polimaster PM1603A/B PM1604A/B
| Detector: | Geiger-Mueller-counter tube |
| Indicating range for dose rate equivalent: PM 1603A/PM 1604A: PM 1603B/PM 1604B: |
0.01 µSv/h .. 6.5Sv/h 0.01µSv/h .. 13Sv/h |
| Adjustment range for threshold of dose and dose rate (each two thresholds): (Adjustment step:) |
Whole measurement range smallest unit |
| Max. realtive intrinsic error for dose rate in the range of: 1 µSv/h .. 5 Sv/h (PM1603A/PM1604A) 1 µSv/h .. 10 Sv/h(PM1603B/PM1604B) |
±(15 + 0.02/H + 0.003H) ±(15 + 0.02/H + 0.002H) H means dose rate equivalent |
| Indicating range for dose equivalent: Measurement range for dose equivalent: |
0.01 µSv .. 9.99 Sv 0.1 µSv .. 9.99 Sv |
| Max. relative intrinsic error for dose equivalent: | ±15% |
| Energy ranges: for PM1603A / PM1603B for PM1604A / PM1604B |
0.048 .. 3MeV 0.048 .. 6 MeV |
| Energy dependency for 0,662 MeV (Cs-137): for PM1603A / PM1603B for PM1604A / PM1604B |
< ±30% (0.048 .. 3MeV) < ±30% (0.048 .. 3MeV) < ±50% (3.0 .. 6MeV) |
| Able to operate within 5 minutes after short gamma irradiation at the max. allowed dose rate of: for PM 1603A/PM1604A for PM 1603B/PM1604B |
50Sv/h 100 Sv/h |
| Additional functions: | Alarm clock, timer, stop watch, calendar |
| Calendar mode: | Day, date, month, year |
| Range for adjustable time spans for timer: Range of the measureable time spans for the stop watch: |
1s .. 23h 59min 59s 0.1s .. 23h 59min 59s |
| Accuracy for clock at standard conditions: | ±5s/day |
| Power supply: | 1 Lithium-cell type CR2032 |
| Battery lifetime: |
9 months |
| Indication of the running and critical battery discharge: |
Indication on LC-display |
| Allowed operating conditions: Ambient temperature: Rel. humidity at 35°C: Air pressure: |
-30 .. +70°C up to 98% 84 .. 106.7kPa |
| Degree of protection (housing): | IP67 |
| Water proof (at short dipping): | down to 1 meter |
| Dimensions: with clip: |
50 x 56 x 19 mm 50 x 90 x 19 mm |
| Weight (incl. battery): | 85g |